In step
1, the test is just getting started. The pressure should be equal to atmosphere, because the vent valve is open, letting fresh air in. Atmosphere tends to be around
760mm/hg absoluteAbsolute pressure means the literal amount of pressure being applied just by the weight of the earth's atmosphere
, or
0 mm/hg gaugeGauge pressure means the pressure you would read off a guage, which is calibrated to earth's atmospheric pressure. For example, your tire gauge says 0 when it's not in use, even though it's subject to 760mm/hg of absolute pressure.
. This will always vary a bit, because air pressure tends to change by very tiny amounts.
In step
2, the system does a
reference check. The vent valve has a hole that's .02" wide in it. It opens this hole and turns on a vacuum pump, causing the pressure in the canister to drop (remember, vacuum is just another word for
negative pressure). During the reference check, the system is seeing what the pressure/vacuum reading would be
if there were in fact a .02" leak. Remember, 1) a leak would cause a loss of pressure/vacuum and 2) in order for it to pass EPA regulations, there cannot be a leak greater than .02".
Typically, the numbers will be about 20mm/hg lower when this reference check runs, so your absolute pressure should be around 740 and your gauge pressure should be about -20, give or take a little.
In step
3, the system closes both valves (purge and vent) and applies vacuum to the canister and fuel tank together, reading the pressure. If the system is good, the vacuum pump should be able to draw a
deeper vacuum than it did during the reference check. In other words, if the vacuum in step 3 goes
below the level set during the reference check (e.g. the reference vacuum is -20mm and the test vacuum is -25mm), the system is
good.
The graph here resembles what a good evap check should look like.
In step
4, it releases the vacuum and pulls down another reference check just to confirm it hasn't changed. It releases the vacuum to prove that none of the valves are sticking shut, then does the second reference check to confirm that nothing went wrong with the reference hole. On the final graph, you should be able to draw a line connecting the two reference checks straight across, and the vacuum level achieved during the test should be below that line.
In step
5, it releases the vacuum one more time, and the test is done.